5-11
Servicing
Troubleshooting
To check the line-power fuse
1
Locate the line-input connector on the instrument’s rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to pry open the fuse holder door.
C A U T I O N
You must remove the voltage tumbler to change the voltage selector. Rotating
the voltage tumbler while it is in the line module damages the line module.
4
Remove the voltage tumbler, and replace the tumbler so that the desired line
voltage value shows through the small opening in the fuse holder door.
5
The fuse is housed in a small container next to the voltage tumbler. Insert the
tip of a screwdriver on the side of the container and gently pull outward to
remove the container.
A spare fuse is stored below the line fuse.
6
Verify that the value of the line-voltage fuse is correct.
• For 100/120V operation, use an IEC 127 5
×
20 mm, 0.315 A, 250 V, Agilent
part number 2110-0449.
• For 220/240V operation, use an IEC 127 5
×
20 mm, 0.16 A, 250 V, Agilent part
number 2110-0448.
W A R N I N G
For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 0.315A/250V for 100/120V
operation and 0.16A/250V for 220/240V operation). The use of other
fuses or materials is prohibited.
Summary of Contents for 83430A
Page 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Page 9: ......
Page 11: ......
Page 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Page 25: ......
Page 31: ...2 6 Making Measurements Using the Agilent 83430A...
Page 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Page 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Page 51: ......
Page 63: ...5 6 Servicing General Information...
Page 69: ...5 12 Servicing Troubleshooting...
Page 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Page 103: ......