2-10
Making Measurements
Cleaning Connections for Accurate Measurements
tions take repeatability uncertainty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be re-
quired? For example, many DFB lasers cannot operate with reflections from
connectors. Often as much as 90 dB isolation is needed.
Figure 2-1. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most
popular connector for fiber-optic applications. While not the highest perform-
ing connector, it represents a good compromise between performance, reli-
ability, and cost. If properly maintained and cleaned, this connector can
withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most
connectors, including the FC/PC style, can deliver. These instruments cannot
tolerate connectors with the large non-concentricities of the fiber common
with ceramic style ferrules. When tighter alignment is required, Agilent
Technologies instruments typically use a connector such as the Diamond
HMS-10, which has concentric tolerances within a few tenths of a micron. Agi-
lent Technologies then uses a special universal adapter, which allows other
cable types to mate with this precision connector. See
.
Summary of Contents for 83430A
Page 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Page 9: ......
Page 11: ......
Page 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Page 25: ......
Page 31: ...2 6 Making Measurements Using the Agilent 83430A...
Page 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Page 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Page 51: ......
Page 63: ...5 6 Servicing General Information...
Page 69: ...5 12 Servicing Troubleshooting...
Page 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Page 103: ......