7
General Maintenance
To Clean an EI XTR, SS, or Inert Source
204
5977B Series MSD Operating Manual
To Clean an EI XTR, SS, or Inert Source
Materials needed
•
Abrasive paper (5061-5896)
•
Alumina abrasive powder (8660-0791)
•
Aluminum foil, clean
•
Cloths, clean (05980-60051)
•
Cotton swabs (5080-5400)
•
Glass beakers, 500 mL
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Solvents
•
Acetone, reagent grade
•
Methanol, reagent grade
•
Methylene chloride, reagent grade
•
Ultrasonic bath
Preparation
1
Disassemble the ion source. (See
“To Disassemble the EI SS or EI Inert
“To Disassemble the EI XTR Source”
2
Collect the following parts to be cleaned for an EI SS or Inert source: (See
•
Repeller
•
Interface socket
•
Source body
•
Repeller block insert
•
Drawout plate
•
Drawout cylinder
•
Ion focus lens
•
Entrance lens
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......