Analyzing Samples
44
Agilent 5500 Series FTIR Operation Manual
Figure 13.
Applying a sample to the DialPath bottom window
Although it is safe to run a variety of liquid samples, including,
aqueous solutions or even thick pastes such as grease, the DialPath is
not to be used with any solid or powder samples, such as tablets. Use
of the DialPath with solid samples will damage the ZnSe windows or
modify the pre-set transmission pathlength.
CAUTION
The top and bottom windows are made from ZnSe. ZnSe can be damaged by
samples with a pH below 4 and above 9. Only measure samples with a pH
between 4 and 9.
CAUTION
ZnSe windows can be easily scratched by hard or abrasive samples. Avoid use of
samples that may scratch the surface of the windows.
CAUTION
Do not break the spectrometer seal and attempt to clean interior surfaces.
Breaking the seal will void the warranty.