Getting Started
Agilent 5500 Series FTIR Operation Manual
19
Table 2.
Diagnostic values
Value
Optimal (green)
Marginal (yellow)
Critical (red)
Comment
Energy
(centerburst)
26,000–21,000 >30,000or
< 15,000
> 31,000 or
< 10,000
Indicates the overall
alignment and proper gain
adjustment for the system.
Battery life
AC or
>30 min
30–5 min
<5 min
Source 1.9 A
>2.2
or
<1.6
>2.5 or
<1.0
Indicates a problem with
the source control voltage
or that the source has
burnt out. Both voltage and
amperage are displayed;
however, amperage
provides a sufficient
diagnostic.
Laser signal 15,000–3,000
>15,000 or
<3,000
>17,000 or
< 2,000
Can verify gross alignment
errors even if the
reflectance cap is not in
place.
Detector
temperature
35–44 <35
or >44
<30
or >48
Indicates a problem with
the cooling circuitry or an
ambient temperature over
the specified range.
CPU
temperature
10–75
>75
>80
Indicates the ambient
temperature over the
specified range.