46
Chapter 3: Service
Calibration Testing Procedures
Figure 14
Bandwidth Test Setup
7
Adjust the signal generator power output for exactly -6.0 dBm as read
on the input power meter.
8
Note the power level reading on the output power meter. 50 MHz
power level _______________ dBm.
The output power level will be approximately -26 dBm. This corresponds to
the 10:1 division ratio of the probe.
9
Change the signal generator frequency to 2.5 GHz.
10
Set the power meter calibration factors to the 2.5 GHz value on the
power sensors.
11
Re-level the signal generator output power for a -6.0 dBm reading on
the input power meter.
12
Note the power level reading on the output power meter. 2.5 GHz
power level _______________ dBm
13
Subtract the reading in step 8 from the reading in step 12 and record
the result in the “Calibration Test Record” on page 47.
The difference should be 3.0 dB .
If the test fails
Go to the troubleshooting section in this chapter.