Chapter 3: Service
Making Adjustments
Use the figure below to locate the appropriate measurement points in the
probe power housing.
Figure 15
Probe Power Box Adjustment Locator
9
Connect the DVM to measure the voltage between ground (the "N"
connector) and the center of the adjustment pot R13.
10
Adjust R13 for a DVM reading of 0.0 V 25 mV.
11
Connect the DVM to measure the voltage between ground and J1 pin 4
(green wire of the cable connector) and record the voltage reading.
_______________mV
This voltage is typically less than 5 mV. Measure it with 10 mV resolution.
12
Multiply the reading in the previous step by -2.3. Observe the signs.
_______________mV
13
Connect the DVM to measure the voltage between ground and the
output of the probe at the 50-
Ω
feedthrough.
14
Adjust R13 for a reading the same as the result obtained in step 12,
within 100 mV.
15
Disconnect the equipment and reassemble the probe. .
If the adjustment cannot be made, see the "Troubleshooting and Repair"
section in this chapter.
J1, Pin 4
(Green Wire)
R13, Center