39
Chapter 3: Service
General Information
Table 6
Service Strategy
The 54701A Active Probe is a high-frequency instrument with many critical
relationships between parts. For example, the frequency response of the
amplifier on the hybrid is trimmed to match the output coaxial cable. As a
result, to return the probe to optimum performance requires factory repair. If
the probe is under warranty, normal warranty services apply. If the probe is
not under warranty, a failed probe can be exchanged for a reconditioned one
at a nominal cost.
See Also
"Troubleshooting and Repair" for further information.
Recommended Test Equipment
Equipment
Required
Critical Specifications
Recommended
Model/Part
Use
Signal Generator
50 MHz to 2.5 GHz
8663A
C
Power Meters (2)
or one Dual-
Channel
50 MHz to 2.5 GHz,
±3% accuracy
436A (2),
437A (2), or
438A (1)
C
Power Sensor (2
50 MHz to 2.5 GHz, 300 mW
8482A
C
Power Splitter
dc to 2.5 GHz,
≤
0.2 dB output tracking,
Type-N
11667A
C
Power Supply
Power and control for probe under test 1143A
C,A,T
DVM
Resistance ±0.1,% Volts and ohms
±0.01%
3458A
C,A,T
Power Supply
5 Vdc
6114A
C
Adapter/termination N(f)-to-probe, 50
Ω
11880A
C
Adapter
N(f-f), 50
Ω
1250-1472
C
Adapter
N(f)-to-BNC(m), 50
Ω
1250-0077
C,A,T
Termination
50
Ω
, BNC feed-through
10100C
C,A,T
Adapter
BNC (f) to banana (m
1251-2277
C
C = CalibrationTests, A = Adjustments, T = Troubleshooting