Chapter 3: Service
Calibration Testing Procedures
Table 7
Calibration Test Record
54701A Active Probe
Tested by_________________________
Serial No. ______________________________
Work Order No.____________________
Recommended Test Interval - 1 Year/2000 hours
Date____________________
Recommended next testing_________________
Temperature_____________
Test
Limits
Results
Input Resistance
100 k
Ω
±1%, 99.0 k
Ω
to 101.0 k
Ω
_____________
dc Gain Accuracy
0.10 ± 0.5%, 0.09950 to 0.10050
_____________
Bandwidth
down less than 3 dB at 2.5 GHz
_____________