DC Function Offset Verification
This test verifies the DC offset and DC output specifications.
Set the function generator for each output indicated in the table below.
Use a DMM to measure the function generator dcV output. Compare the
measured results to the test limits shown in the table. This is a HIGH Z
output termination test.
Agilent 33120A
Measurement
Function
OUT TERM
1
Ampl
Nominal
Error
Q
DC Volts
HIGH Z
10.0 Vdc
10.0 Vdc
0.20 Vdc
DC Volts
HIGH Z
-10.0 Vdc
-10.0 Vdc
0.20 Vdc
AC Amplitude Verification
This procedure is used to check the output amplitude calibration of the
function generator. Verification checks are performed to check the
accuracy of the pre-attenuator and post attenuator.
Make sure you have
read “Test Considerations” on page 53.
Set the function generator for each output indicated in the table on the
next page. Use a DMM to measure the ACrms output voltage of the
function generator. Compare the measured results to the test limits
shown in the table. This is a HIGH Z output termination test.
1
Output termination set using front panel controls. HIGH Z assumes no load on
output. 50
W
assumes a 50
W
0.1
W
load on output.
4
Chapter 4 Calibration Procedures
DC Function Offset Verification
57
Summary of Contents for 33120A
Page 2: ......
Page 10: ...8 ...
Page 14: ...Contents 12 ...
Page 15: ...1 Specifications 1 ...
Page 20: ...18 ...
Page 21: ...2 Quick Start 2 ...
Page 37: ...3 Front Panel Menu Operation 3 ...
Page 51: ...4 Calibration Procedures 4 ...
Page 85: ...5 Theory of Operation 5 ...
Page 103: ...6 Service 6 ...
Page 114: ...112 ...
Page 115: ...7 Replaceable Parts 7 ...
Page 129: ...8 Schematics 8 ...
Page 145: ...TOP SIDE 33120 66502 REV C 33120 66502 Component Locator Diagram 143 ...