To set the output amplitude
At power-on, the function generator outputs a sine wave with an
amplitude of 100 mV peak-to-peak (into a 50
W
termination).
The following steps show you how to change the amplitude to 50 mVrms.
Ampl
1 Enable the
amplitude modify
mode.
The displayed amplitude is either the power-on value or the previous
amplitude selected. When you change functions, the same amplitude is
used if the present value is valid for the new function.
1
0
0.0 mVPP
Enter Number
2 Enter the magnitude of the desired amplitude.
1
5
0
Notice that the
Num
annunciator turns on and “
ENTER NUM
” flashes on
the display, indicating that the number mode is enabled.
50
To cancel the number mode, press
Shift Cancel
.
Shift
3 Set the units to the desired value.
¿
kHz
m Vrms
The units are selected using the arrow keys on the right side of the front
panel. As soon as you select the units, the function generator outputs the
waveform with the displayed amplitude.
To turn off the flashing digit,
move the cursor to the left of the display using the arrow keys.
5
0
.00 mVRMS
1
You can also use the knob and arrow keys to enter a number.
Chapter 2 Quick Start
To set the output amplitude
26
Summary of Contents for 33120A
Page 2: ......
Page 10: ...8 ...
Page 14: ...Contents 12 ...
Page 15: ...1 Specifications 1 ...
Page 20: ...18 ...
Page 21: ...2 Quick Start 2 ...
Page 37: ...3 Front Panel Menu Operation 3 ...
Page 51: ...4 Calibration Procedures 4 ...
Page 85: ...5 Theory of Operation 5 ...
Page 103: ...6 Service 6 ...
Page 114: ...112 ...
Page 115: ...7 Replaceable Parts 7 ...
Page 129: ...8 Schematics 8 ...
Page 145: ...TOP SIDE 33120 66502 REV C 33120 66502 Component Locator Diagram 143 ...