To unsecure the function generator for calibration
The function generator can use a calibration security code to prevent
unauthorized or accidental calibration. This procedure shows you how to
unsecure the function generator for calibration.
Shift
1 Turn on the menu.
Menu On/Off
A
: MOD MENU
<
2 Move across to the CAL MENU choice on this level.
F:
CAL MENU
¿
3 Move down a level to the
SECURED
command.
1
: SECURED
If the display shows UNSECURED, you do not need to perform this
procedure to execute a calibration.
3
Chapter 3 Front-Panel Menu Operation
To unsecure the function generator for calibration
47