3200D Dissolved Oxygen Meter Operating Guide
3
Troubleshooting and Maintenance
Troubleshooting the 3200D Dissolved Oxygen Meter
lists common problems, possible causes, and how to
resolve them.
Note that improper operation of the DO probe causes abnormal
readings. During measurement, soak the probe tip completely in
solution.
Table 3
3200D Troubleshooting
Number
Failure mode
Failure cause
Solutions
1
No display after meter starts
up. The display is not lit up.
1
Power adaptor not installed correctly.
2
The power supply does not meet with the
requirements.
3
The power adaptor is damaged.
4
The power socket has poor contact.
5
The LCD is damaged (After start up for a
few minutes, the user can hear a buzzing
sound after pressing
[On/Off]
key, but
there is no display on LCD.)
1
Reconnect.
2
Use the required power
supply.
3
Replace the power adaptor.
4
Ensure there is a good
contact.
5
Contact Customer Service.
2
Display is dim.
Check the power adapter.
3
No buzzing when pressing
key
1
You may have pressed invalid keys under
the current setup.
2
The buzzer has been damaged.
The buzzer only sounds when
you press a valid key.
4
No response when pressing
key
1
You may have pressed invalid keys under
the current setup.
2
The key has been damaged.
Press the valid key for
operation.
Summary of Contents for 3200D
Page 1: ...Agilent Technologies Agilent 3200D Dissolved Oxygen Meter Operating Guide...
Page 6: ...6 3200D Dissolved Oxygen Meter Operating Guide Contents...
Page 52: ...52 3200D Dissolved Oxygen Meter Operating Guide 2 Operation...
Page 63: ...Agilent Technologies Agilent 3200D...
Page 67: ...67 Agilent Technologies Agilent 3200D 1 3200D 68 3200D 69 72 73...
Page 68: ...68 3200D 1 Agilent G4389A D6111 DO Probe 5190 3997 5185 8389...
Page 69: ...1 3200D 69 3200D 3200D 1 2 3 4 5 6 7 8...
Page 70: ...70 3200D 1 5190 3997...
Page 71: ...1 3200D 71 100 240 VAC 1 A 50 60 Hz DC501...
Page 72: ...72 3200D 1 G4390A USB...
Page 73: ...1 3200D 73 3200D...
Page 74: ...74 3200D 1...
Page 75: ...75 Agilent Technologies Agilent 3200D 2 3200D 76 83 101 3200D...
Page 77: ...2 3200D 77 5 3200D 3200D GLP GLP 200 USB PC PC bug...
Page 79: ...2 3200D 79 3 3200D 3200D...
Page 80: ...80 3200D 2 3200D 1 2 3200D 1 2 3 USB 4...
Page 81: ...2 3200D 81 3200D On Off 3200D 15 1 2 3 4 5 6 7 8 9 0 1 1 2 4 8 6 2 4 8 6 3 3 5 5 7 7...
Page 82: ...82 3200D 2 9 9 0 0...
Page 83: ...2 3200D 83 3200D 3200D 3200D 3200D 3200D 3200D 3200D On Off...
Page 84: ...84 3200D 2 1 2 GLP 1 2...
Page 85: ...2 3200D 85 3 4 3 4...
Page 86: ...86 3200D 2 5 4 6 4 6 30 USB PC 1 99 10 5...
Page 87: ...2 3200D 87 6 1 0 1 0 s 1 200 GLP 7 h 6...
Page 88: ...88 3200D 2 000 200 7 200 201 201...
Page 89: ...2 3200D 89 8 9 101 3kPa 8 9...
Page 90: ...90 3200D 2 10 0 0g L 10 480min 0 nA 1000 nA 25 0 C 101 3 kPa 0 0 g L 10 min 000 1 0 10...
Page 91: ...2 3200D 91 5 s 0 1 2 3 5190 0547 4 5 60 1 1 25 20 80 cm 2 2...
Page 92: ...92 3200D 2 11 5 12 11 12...
Page 93: ...2 3200D 93 13 14 13...
Page 94: ...94 3200D 2 14...
Page 95: ...2 3200D 95 200 200 USB PC PC...
Page 98: ...98 3200D 2 17 USB PC MODEL 3200D DISSOLVED OXYGEN DETERMINATOR VER 1 00 PRINT TIME 16 17...
Page 100: ...100 3200D 2 5 6 7 60...
Page 101: ...2 3200D 101 3200D 3200D G4390A PC DVD 1 On Off 2 USB PC 3 PC 1 On Off 2 PC 3 PC...
Page 102: ...102 3200D 2...
Page 103: ...103 Agilent Technologies Agilent 3200D 3 3200D 104 105 106...
Page 104: ...104 3200D 3 1 1 2 3 4 5 1 2 3 4 5 2 1 2 1 2 3 1 2 1 2 4 1 2 1 2...
Page 105: ...3 3200D 105 ATC 5185 8390 680 720nA 0 5nA 300 1000nA ATC 5185 8390 49 0 51 0...
Page 106: ...106 3200D 3 18...
Page 107: ...3 3200D 107 19...
Page 109: ...109 Agilent Technologies Agilent 3200D 4 110 111...
Page 110: ...110 3200D 4 1 2 3 4 5 6 60 1 1 25...
Page 111: ...4 3200D 111...
Page 112: ...112 3200D 4...
Page 113: ...113 Agilent Technologies Agilent 3200D 5 3200D 3200D...
Page 115: ......