Displaying Data
2
Agilent 3000 Series Oscilloscopes User’s Guide
39
To adjust the horizontal position
•
Turn the horizontal position knob to change the position of the trigger
point relative to the center of the screen.
The position knob adjusts the horizontal position of all channels, math
functions, and reference waveforms.
There is also a menu item to return the trigger position to the center of
the screen (see
“To reset the trigger offset”
on page 42).
To display the delayed sweep time base
The delayed sweep time base magnifies a portion of the original waveform
display (now on the top half of the screen) and displays it in a zoomed
time base on the bottom half of the screen.
1
Press
Main/Delayed
.
2
In the Main/Delayed menu, select
Delayed
to toggle the delayed sweep
time base “ON” or “OFF”.
3
When the delayed sweep time base is “ON”:
•
The top half of the display shows the original waveform and the
portion being magnified.
•
The horizontal scale knob changes the magnification (widens or
narrows the area of magnification).
•
The horizontal position knob moves the area of magnification
forward and backward on the original waveform.
•
The bottom half of the display shows the magnified data in the
delayed sweep time base.
N O T E
The delayed sweep time base setting cannot be set slower than the original waveform’s
time base setting.
Summary of Contents for 3000 Series
Page 1: ...A Agilent 3000 Series Oscilloscopes User s and Service Guide ...
Page 12: ...12 Agilent 3000 Series Oscilloscopes User s Guide Contents ...
Page 16: ...16 Agilent 3000 Series Oscilloscopes User s Guide Tables ...
Page 34: ...1 Getting Started 34 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 126: ...7 Specifications and Characteristics 126 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 150: ...A Safety Notices 150 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 155: ......