8
Service
142
Agilent 3000 Series Oscilloscopes User’s Guide
18
Calculate the reference gain as follows:
Record this value in the Calculated Gain @ 1 MHz column of the
“Performance Test Record”
on page 144.
19
Change the signal generator frequency to the value for the model being
tested as shown in the table below.
20
Change the oscilloscope time base to the value for the model being tests
as shown in the table above.
21
Using the Vpp reading, calculate the Vrms value using the following
expression and record it in the
“Performance Test Record”
on page 144:
For example, if Vpp = 1.24 V:
22
Using the power meter reading, convert this measurement to Volts RMS
using the expression and record it in the
“Performance Test Record”
on
page 144:
For example, if Pmeas = 3.65 mW:
Table 20
Oscilloscope Models and Signal Generator Frequency
Setting
Model
DSO3062A
DSO3102A
DSO3152A
DSO3202A
Frequency
60 MHz
100 MHz
150 MHz
200 MHz
Time Base
10 ns/div
5 ns/div
5 ns/div
2 ns/div
Gain
1
MHz
Vout
1
MHz
Vin
1
MHz
----------------------------
=
Vout
max
Vpp
max
2 2
-------------------
=
Vout
max
1.05
2 2
----------
1.05
2.828
-------------
371 mV
=
=
=
Vin
max
P
meas
50
Ω
×
=
Summary of Contents for 3000 Series
Page 1: ...A Agilent 3000 Series Oscilloscopes User s and Service Guide ...
Page 12: ...12 Agilent 3000 Series Oscilloscopes User s Guide Contents ...
Page 16: ...16 Agilent 3000 Series Oscilloscopes User s Guide Tables ...
Page 34: ...1 Getting Started 34 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 126: ...7 Specifications and Characteristics 126 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 150: ...A Safety Notices 150 Agilent 3000 Series Oscilloscopes User s Guide ...
Page 155: ......