Service
8
Agilent 3000 Series Oscilloscopes User’s Guide
129
Before Testing Performance
Equipment Required
Calibration
1
Push the
Utility
button on the front panel.
2
Select
Self-Cal
menu item in the Utility menu.
3
Follow the on- screen instructions.
N O T E
Let the oscilloscope warm up before testing.
The oscilloscope under test must be warmed up (with the oscilloscope application running)
for at least 30 minutes prior to the start of any performance test.
Table 15
Equipment Required for Performance Verification Testing
Description
Critical Specifications
Recommended
Model/Part Numbers
Digital Multimeter
DC voltage measurement accuracy better
than ±0.1% of reading
Agilent 34401A
Cable Assembly
50
Ω
characteristic impedance
Agilent 54855-61620
Cable Assembly
RS-232 (f)(f)
Agilent 34398A
Adapter
BNC Barrel (f)(f)
Agilent 1250-0080
Adapter
BNC shorting cap
Agilent 1250-0929
Adapter
Precision BNC (2)
Agilent 54855-67604
Adapter
BNC (f) to dual banana
Agilent 1251-2277