58
2200 TapeStation
6
Troubleshooting
Troubleshooting 2200 TapeStation ScreenPlex Images
The bands look
smudged – especially
towards the top of
the lane.
The sample has not
been mixed properly
with the loading
buffer.
Always vortex samples for 5 seconds then spin
down before placing in the 2200 TapeStation
All of the sample
and/or ladder bands
appear very faint.
The lane has not
loaded optimally.
Always spin down samples before placing in the
2200 TapeStation to remove any bubbles.
Very rarely, the 2200 TapeStation does not load
effectively - run the sample again.
Table 8
Overview troubleshooting 2200 TapeStation images
Image
Issue
Probable cause
Solution