2200 TapeStation
53
Troubleshooting
6
ScreenTape
ScreenTape
Table 5
ScreenTape
Issue
Probable cause
Solution
A bubble has
appeared in the
buffer chamber of the
tape.
This may occur over
time.
Holding the tape with the ScreenTape label facing
you, flick the bottom of the tape. This will move
any bubbles to the top of the chamber.