40
11713B/C Operating and Service Manual
3
Specifications
Environmental Specifications
Agilent 11713B/C attenuator/switch drivers are designed to fully
comply with Agilent Technologies’ product operating environmental
specifications shows in table below.
Table 9
11713B/C Environmental Specifications
Temperature
•
Operating
•
Storage
0
°
C to +50
°
C
-40
°
C to +70
°
C
Humidity
•
Operating
•
Storage
•
Condensing
95% RH at 40
°
C, 5 days cyclic
90% RH at 65
°
C, 24 hours
95% RH at 40
°
C, 5 hours (condensation 15 minutes)
Shock
•
End-user handling
•
Bench Handling
•
Functional
•
Transportation
Half-sine: 2-3 ms duration, 60 in/s (1.6 ms) delta-V
Half-sine: 11 ms duration, 30 grms
Trapezoital: 18-22 ms duration, 337 in/s (8.56 ms) delta-V
Per MIL-PRF-28800F
Vibration
•
Operating
•
Survival
Random: 0.21G rms, 5 to 500 Hz, 10 min/axis
Random: 2.09G rms, 5 to 500 Hz, 10 min/axis
Swept-sine: 0.5 G rms, 5 to 500 Hz, 10 min/axis
Altitude
•
Operating
•
Non-operating
< 2,000 meters (6,540 feet)
< 15,300 meters (50,000 feet)
Summary of Contents for 11713B
Page 1: ...Agilent Technologies Agilent 11713B C Attenuator Switch Drivers Operating and Service Manual ...
Page 12: ...12 11713B C Operating and Service Manual ...
Page 24: ...24 11713B C Operating and Service Manual 1 Introduction ...
Page 43: ...Specifications 3 11713B C Operating and Service Manual 43 Declaration of Conformity ...
Page 44: ...44 11713B C Operating and Service Manual 3 Specifications ...
Page 64: ...64 11713B C Operating and Service Manual 5 Local Operations ...
Page 88: ...88 11713B C Operating and Service Manual 7 Remote Interface Configurations ...
Page 104: ...104 11713B C Operating and Service Manual 8 Replacing Major Assemblies ...
Page 114: ...114 11713B C Operating and Service Manual 9 Servicing the Attenuator Switch Driver ...