![Agilent Technologies 1168/9A User Manual Download Page 212](http://html.mh-extra.com/html/agilent-technologies/1168-9a/1168-9a_user-manual_2867526212.webp)
212
1168A/9A Probe’s User’s Guide
7
Performance Verification
To Test Bandwidth
37
Select
[display]
key then
data->memory
screen key.
38
You have now saved V
in
waveform into the VNA's memory
for future use.
Measuring Vout Response
39
Disconnect the
PORT 2
cable from PV/DS test board and
attach to probe output on the AutoProbe Adapter.
40
Connect the Calibration Standard load to PV/DS test board
(non- pincher side) as shown in
.
Figure 176 Setup for Measuring V
out
41
Press
[scale reference]
key on the VNA.
42
Set reference value to –10.8 dB.
43
The display on screen is V
out
and it should be similar to
that shown in
.
Summary of Contents for 1168/9A
Page 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1168 9A Probes User s Guide Contents ...
Page 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Page 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Page 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Page 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Page 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Page 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Page 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...