1168/9A Probes User’s Guide
4
Calibrating Probes
Calibration for Hand-Held Browser Probe Heads
7
In the Channel Setup dialog box select the Probes... button.
8
In the Probe Setup dialog box select the Calibrate Probe...
button.
9
In the Probe Cal dialog box select the Calibrated
Atten/Offset radio button.
10
Select the Start Atten/Offset Calibration... button and follow
the on- screen instructions for the vertical calibration
procedure.
11
Once the vertical calibration has successfully completed,
select the Calibrated Skew... button.
12
Select the Start Skew Calibration... button and follow the
on- screen instructions for the skew calibration.
Figure 91
Placing the Probe on the Fixture
Summary of Contents for 1168/9A
Page 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1168 9A Probes User s Guide Contents ...
Page 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Page 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Page 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Page 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Page 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Page 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Page 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...