Performance Verification
7
To Test Bandwidth
1168A/9A Probe’s User’s Guide
207
Figure 171 PV/DS Test Board Connected to VNA
c
Select the
open
screen key under the
Forward
group.
d
Wait until the VNA beeps indicating that it has completed
the task.
e
Connect short end of Calibration Standard to the
non- pincher side of the PV/DS test board.
f
Select
short
screen key under the
Forward
group.
g
Wait until the VNA beeps indicating that it has completed
the task.
h
Connect load end of Calibration Standard to the
non- pincher side of the PV/DS test board.
i
Select the
loads
screen key under the
Forward
group.
j
Press
broadband
screen key selection.
k
Wait until the VNA beeps indicating that it has completed
the task.
l
Press the
done loads
screen key.
m
You have just calibrated one side of the reference plane.
Summary of Contents for 1168/9A
Page 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1168 9A Probes User s Guide Contents ...
Page 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Page 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Page 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Page 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Page 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Page 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Page 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...