Calibrating Probes
4
N2887A & N2888A Calibration and Deskew Procedure
1168/9A Probes User’s Guide
135
■
For the N2887A probe head, use the
connection with the white molded
strain relief and the yellow primary
banding.
■
For the N2888A probe head, use the
connection with the black molded
strain relief and yellow primary
banding.
NOTE
To minimize the wear and tear on the probe head, the probe head should
be placed on a support to relieve the strain on the probe head cables.
NOTE
Please note that you are only deskewing one channel on the InfiniiMax
Soft Touch probe head in this procedure. The other channels are nominally
identical (< 15 ps). The major source of skew when using multiple
InfiniiMax amplifiers with a single InfiniiMax Soft Touch probe head is the
variations from one amplifier to another. This skew can be addressed by
repeating the procedure outlined above for each InfiniiMax amplifier that
will be used.
NOTE
Once you have the probe head connected correctly, perform a normal
probe calibration as described in the Infiniium help system available on the
oscilloscope.
Summary of Contents for 1168/9A
Page 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1168 9A Probes User s Guide Contents ...
Page 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Page 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Page 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Page 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Page 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Page 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Page 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...