Positioning Parameters
On systems with automatic positioning, the target position is automatically
set, based on the selected exposure.
To modify the target position, you can:
• scroll through the preconfigured set of target positions by clicking the
automatic position selection buttons. Each target position has a reference
number. A reference image is displayed. The target positioning parameters
are set.
• increase or decrease the positioning parameters step by step: use the UP
and DOWN arrows. The values increase or decrease step by step each time
the corresponding button is touched.
Figure 20: Positioning parameters
1.
Automatic position selection buttons
2.
Actual positioning parameters
3.
Target positioning parameters
The selected modality position defines which positioning parameters are
available.
Note
:
Depending on the system configuration, the target
positioning parameters may not be editable.
To move the X-ray system to the target position, push and hold the MOVE
button on the X-ray system or on the remote control. When the X-ray system
reaches the target position, the values for the actual position parameters
match the values for the target position parameters and the “On target” status
is displayed in the device status frame.
DX-D Software Console | X-Ray Modality Controls Window |
45
0189D EN 20180329 1421