Sharing the DR Detector between NX Workstations
To share a DR Detector between NX Workstations, the DR Detector must be
configured on each of the workstations. An IR Data Communication Unit is
connected to each of the workstations.
Note
:
The IR Data Communication Unit is configured to be
connected to a specific USB port. Do not connect it to another USB
port.
The DR Detector is set up to communicate to a specific NX workstation. The
procedure of registering the DR Detector to another NX workstation switches
the availability of the DR Detector between NX workstations.
There are two workflows for registering the DR Detector. Which workflow is
used, is configured on the NX workstation during installation.
• Using automatic registration.
The registration is initiated by holding the detector with its IR data port
close to the IR Data Communication Unit that is connected to the NX
workstation.
• Using the
DR 10s DR 14s Registration Tool
The registration is initiated by running the tool on the NX workstation.
Topics:
•
Registering the DR Detector to an NX Workstation using automatic
registration
•
Registering the DR Detector to an NX Workstation using the DR 10s DR
14s Registration Tool