![AGFA DR 10s User Manual Download Page 109](http://html.mh-extra.com/html/agfa/dr-10s/dr-10s_user-manual_2866637109.webp)
Recommended protective dis-
tance:
Conducted high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-6
3 V
eff
150 kHz
to 80 MHz
3 V
eff
150 kHz
to 80
MHz
d = 1.2
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2
80 MHz to 800
MHz
d = 2.3
800 MHz to 2.5
GHz
With P as the rated power of the
transmitter in watts (W) in ac-
cordance with the manufacturer
information on the transmitter
and d as the recommended pro-
tective distance in metres (m).
The field strength of stationary
radio transmitters is lower than
the level of the agreement at all
frequencies in accordance with
an on-site investigation.
Disruptions are possible near
devices that carry the following
symbol:
Note
:
The higher value will apply at 80 MHz and 800 MHz.
DR 10s | Remarks for HF-emission and immunity |
109