Operation Conditions
Temperature and humidity levels during operation:
Table 6: Allowed temperature and humidity levels during operation
Temperature
CR HD5.x detectors, CR MD4.xR and
CR MM3.xR plates: 15°C - 30°C (59°F - 86°F)
CR HM5.x detectors: 20°C - 30°C (68°F - 86°F)
Relative humidity
15 to 75% (recommended 30 to 60%),
IEC 721-3-3: class 3K2
Do not place any excessive loads on the CR detectors and CR plates and
cassettes.
Avoid UV-radiation or direct sunlight on the CR detectors and CR plates and
cassettes.
The storage of CR detectors and CR plates and cassettes shall be protected
against radiation in such a way, that the annual dose rate at the place of
installation will not exceed 1m Sv/a.
Use plastic envelops to protect the CR detectors and CR cassettes against
contact with body-fluids.