Compatibility
The use of CR HM5.x Mammo detectors is restricted by compatibility criteria
specified in the table below, depending on the availability of an IP gain
calibration file.
There are two valid system configurations. If one component is upgraded or
replaced, all other components must be upgraded or replaced to fit the new
configuration.
Table 1: Compatibility of CR HM5.x Mammo detectors with the NX software
Configuration 1
NX software
Version NX 2.0.8500/3.0.8500 or older
CR HM5.x Mammo
detector
Without IP gain calibration file
Configuration 2
NX software
Version NX 2.0.8500/3.0.8500 SU1 or newer. For
plates or detectors that are delivered with an IP gain
calibration file, the IP gain calibration files must be
loaded on the NX workstation.
CR HM5.x Mammo
detector
With or without IP gain calibration file
How to retrieve the version number of the NX software is described in the NX
User Manual.