Magnetic field
at the supply
frequency
(50/60 Hz) in
accordance
with IEC
61000-4-8
30 A/m
30 A/m
Magnetic field at the net-
work frequency should
correspond to the typical
values as they are in a
commercial and clinical
environment.
Tests of Resist-
ance to Disrup-
tion
IEC 60601
Test Level
Level of
Agree-
ment
Electromagnetic Environment
Conducted high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-6
3 V
eff
150
kHz to 80
MHz
6 V
eff
in
the ISM
bands 150
KHz to 80
MHz
3 V
eff
150
kHz to 80
MHz
6 V
eff
in
the ISM
bands 150
KHz to 80
MHz
The electromagnetic field
strength of a stationary RF
transmitter determined by an
electromagnetic test survey
must be less than the compli-
ance level of each frequency
range.
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with
IEC 61000-4-3
3 V/m
80 MHz to
2.7 GHz
3 V/m
80 MHz
to 2.7
GHz
Interference may occur in the vi-
cinity of equipment marked
with the symbol:
Note
:
The higher value will apply at 80 MHz and 800 MHz.
Note
:
These Guidelines may not apply to all situations. The dis-
persion of electromagnetic waves is influenced by absorption
and reflections from buildings, objects and people.
DX-D 45C, DX-D 45G | Remarks for HF-emission and immunity |
117