Identifying problems
Please refer to the details of following symptoms or error messages. If the
problem persists, turn off the detector and consult your sales representative or
local dealer.
WARNING:
Improper changes, additions, maintenance or repair of the
system can lead to personal injury, electrical shock and damage
to the equipment. Safety is only guaranteed when changes,
additions, maintenance or repairs are carried out by an Agfa
certified field service engineer. A non certified engineer
performing a modification or service intervention on a medical
device, acts on his own responsibility and makes the warranty
void.
Symptom
Cause
Remedy
The detector will
not turn on.
The battery is not at-
tached.
Attach the battery.
The battery pack is not
charged.
Fully charge the battery pack.
The battery pack is bro-
ken.
Replace the battery pack.
The status indi-
cator of the Sys-
tem Control Unit
does not light
up.
The power cord is un-
plugged from the AC
outlet.
Connect the plug to the AC out-
let firmly. If it still does not
work, replace the System Con-
trol Unit.
The status indi-
cator of the Sys-
tem Control Unit
does not light up
in green.
A hardware error has
occurred.
Turn off the System Control
Unit and turn it on again. If it
still does not work, replace the
System Control Unit.
The green status
indicator lights
up and the or-
ange and blue
status indicators
are flashing.
An error has occurred
during registration of
the DR Detector.
Check the network connection
on the System Control Unit.
Check the network configura-
tion of the workstation.
An error has occurred
during data communi-
cation.
Check if the System Control
Unit is turned on.
DX-D 45C, DX-D 45G | Problem solving |
103
0292B EN 20190401 1627