Step 5: execute the exposure
In the operator room:
Press the exposure button to execute the exposure.
WARNING:
During exposure ionizing radiation is emitted by the X-ray
system. To indicate the presence of ionizing radiation, the
radiation indicator on the control console lights up.
• The actual X-Ray exposure parameters are sent back from the generator to
the NX workstation and are shown in the Image Detail pane.
• The actual X-Ray exposure parameters and the Exposure Index (EI) value
on the NX workstation can be used to monitor the performance of the
Automatic Exposure Control of the X-Ray system.
• A green OK mark appears on all thumbnails for which the exposures are
made and for which exposure settings are sent back to the NX workstation.
96
| Valory (system with floor mounted column) | Basic Workflow
0426A EN 20220804 1858