Figure 40: Navigation button for modality controls
4.
Click the button to start the automated workflow for warming-up of the X-
ray tube.
Figure 41: Button to start the automated workflow for warming-up of the
X-ray tube
A table is displayed with a list of exposures.
1
2
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
xx
1.
Table with list of exposures
2.
Button to cancel the warming-up procedure
Figure 42: List of exposures for warming-up of the X-ray tube
5.
Make sure that the collimator blades are fully closed and that no patient is
present in the room.
To avoid radiation on a DR detector, remove the detector, turn the tube
away from the detector, or cover the detector with a lead apron.
6.
Perform the exposures and wait for the timer icon to finish between
exposures.
The exposure parameters are set automatically.
Valory (system with floor mounted column) | Basic Workflow |
83