1.
X-ray tube
2.
Focal spot label [—]
3.
Significant zone of occupancy.
Minimum area 60x60 cm.
Minimum height above the floor 200 cm.
4.
Dose meter
5.
DR Detector or cassette
6.
Significant zone of occupancy at the left side of the radiographic table
7.
Significant zone of occupancy in front of the radiographic table
8.
Significant zone of occupancy at the right side of the radiographic table
Figure 20: Significant zones of occupancy at the radiographic table
Significant zones of occupancy at the radiographic wall stand
If operator or staff needs to be close to the patient during normal use (e.g.
some pediatric examinations or types of examinations for which the patient
Valory (system with floor mounted column) | Introduction |
37