Wiring the system using individual cables
When fitting an ultrasonic sludge
sensor a junction box should
always be used that allows the
equalisation of the earth potential
outside of the alarm control unit.
When connecting more than 1 sensor it is
recommended that a junction box is used.
If a junction box is not used all cable joints
should be sealed using heat shrink tubing.
When connecting an WGA-ES8 sludge
sensor an external junction box should be
used that allows the shielded cable to be
grounded to earth.
Terminals
K1: 230 VAC, 4 VA
K2: R1 Voltage free relay output (Symbols
on the circuit board are indicated in the
alarm condition)
K2: R2 Voltage free relay output (Symbols
on the circuit board are indicated in the
alarm condition)
K4: Terminal contact for membrane keypad
ribbon cable (not shown)
K3: 1-G Capacitive sensor WGA-ES4
K3: 2-G Thermal sensor WGA-R6-S
K3: 3-G Sludge sensor WGA-ES8
Recommended cable
Power supply: 3 x 1.5 mm²
Separate cable for sensor WGA-ES4 and
WGA-R6-S: 2 x 1 mm²
Separate cable for sensor WGA-ES8:
3 x 1 mm² or 2 x 1 mm² with shielding
230 VAC
Layer
sensor - WGA-
ES4
Damming sensor -
WGA-R6
Junction box
2
1 2
1
2
1
Sludge sensor -
WGA-ES8
+