Mounting and commissioning
– English –
DIT 10
34
results in incorrect measurements.
Close the junction box in such a way that it is water-tight, but not
completely air-tight.
6.5
Connecting the battery
1.
After the pressure sensor and the display unit have been elec-
trically connected, open the housing of the display unit. To do
so, turn the upper part of the housing vis-à-vis the lower part all
the way to the stop and then pull the upper part of the housing
off of the lower part.
2.
Press the mini button (1) and hold it down. Plug the two-pole
battery connector into the two-pole socket (2) on the PCB. Veri-
fy correct polarity when plugging in the connector.
1
Mini button
2
Socket
Fig. 4: PCB
WARNING
Explosion hazard due to short circuit of lithium battery
Do not short-circuit the lithium battery.
Verify correct polarity when plugging the battery connector into
the socket on the PCB in order to avoid a short circuit.
3.
Release the mini button.