22
4. CONTROLLING THE FLOW METER
The flow meter is controlled from a PC running "Viewer BUF-311"
program. The program is available free of charge.
The interface includes several tabs classified by their function. Con-
tents of the tabs are determined by the flow meter's operation mode. Af-
ter the flow meter is powered up, the operation mode is detected auto-
matically.
4.1. Configuring before work
4.1.1. Installing the dedicated software and system requirements
BUF-311 program was designed to control the flow meter and han-
dle the measurement data. Viewer BUF-311 works under Windows 98
(Me, 2000, XP, Vista, 7) operating system.
To install Viewer BUF-311 program, create a directory with the
same name and copy BUF-311.vpr file to this directory. The program
does not require another installation procedure.
4.1.2. Getting started
Connect the serial port of a PC and RS-port of the flow meter
RS232/485 or USB-RS232/485 adapter by a cable. Activate SERVICE
mode (place a jumper on J2 terminal). The list of parameters to be
configured on-site in SERVICE mode is given in Appendix C. Run the
BUF-311.vpr file. Wait for the main window shown in Fig.6 to open.
Fig.6. Main window
Click on the leftmost icon in the command line of the main window.
By doing so, the connection with the flow meter is established. The win-
dow will take the form pictured below (Fig.7).