17
IEC Pass/Fail Analysis
Port 1: Main Menu and Settings
IEC 61300-3-35 defines connector inspection pass/fail criteria .
Pass/Fail criteria depends on:
•
Fiber type (SMF or MMF)
•
Connector end face regions
–
A region: Core
–
B region: Cladding
–
C region: Adhesive (between cladding
and ferrule)
–
D region: Physical contact area
•
Type of end face flaw:
–
Scratches
–
Defects (contamination, particles, chips)
•
Size of flaw
Core
Failing
Defect
Passing
Defect
Contact
Adhesive
Cladding
Shown below is the display when the IEC rule for single-mode PC connector with return loss ≥ 45 dB is
selected . The table identifies the number of scratches or defects of a certain size, which may be allowed
in each of the core, cladding, and contact regions .
Example IEC Rule
Region
Scratch
Defect
A: Core
>0 µm: 0
No scratches >0 µm
allowed
>0 µm: 0
No defects >0 µm
allowed
B: Cladding
>3 µm: 0
No scratches >3 µm
allowed
2-5 µm: 5
Up to 5 defects 2-5 µm
diameter allowed
>5 µm: 0
No defects >3 µm
allowed
C: Adhesive
Blank
(No limitations on
scratches)
Blank
(No limitations on
scratches)
D: Contact
Blank
No limitations on
scratches
0 ≥ 10 µm
No defects ≥ 10 µm
allowed