4.0 Troubleshooting
4.1 Troubleshooting
Guide
Symptom
Possible Cause
Fault isolation/Solution
Gas sensor readings incorrect
Insufficient warm up
Incorrect zero calibration
Incorrect span calibration
Sensor module leaking
Sensor pump failed
Allow the sensors to fully warm up after power
down. This may require 60 minutes or more.
Repeat zero calibration with clean air. Check
media in Gas Treatment module.
Perform span check.
Check for leaks
Measure flow. If pump has failed replace.
NO
2
sensor reading very high
Incorrect zero calibration
O
3
Scrubber failed
Repeat zero calibration with clean air. Check
media in Gas Treatment module.
Challenge the NO
2
sensor with 0.1 ppm of O
3
. If
the response is large (>0.5 ppm) then the
scrubber has failed. Replace.
Gas Sensor readings noisy or
unstable
Leaks
Particulate filter
Leaks dilute the sample stream and can cause
low span readings and incorrect zero readings.
Perform a leak test.
Replace if dirty.
“NR” or “9999” in Data table
(No response)
Occasional NR or “9999” in Data
Table
No Response from Sensor module.
Not connected correctly.
Computer too slow
RF noise in environment.
Check the electrical connector on the sensor
module is firmly connected
Test with a faster computer.
Try to reduce RF noise
COMMUNICATION
No communication over RS232
RS232 cable disconnected
Control Module Failed
Reconnect cable
Replace Control Module
PC Data logging stopped
IQM power blackout interrupted
data logging.
USB to serial hub not working
Close and restart IQM software.
Check USB connectors. Check the Moxa serial
hub is present on the device hardware menu in
the PC. If not reload driver software and re-
install.
No data on SD card
Card not correctly installed in slot.
Data logging interval set too large.
Data card module fault
Turn off IQM and install card in slot correctly.
Set logging interval in configuration to a lower
interval.
Send control module to factory for replacement
module.
RF communication fails
RF modem incorrectly configured
Follow modem manual instructions to
reconfigure RF modem