IQM 60 V5 Pump Revision October 2010
34
* Set gain factor 1.0 for O3 sensor
17A. Acknowledge reply command of setting gain value, 4 bytes:
SENSOR, AQM_ID, SET_GAIN_FACTOR, CHECKSUM
18. Set sensor names for data logging and monitors with displays.
This is 4 bytes command, followed with 115 bytes with sensor code and name ASCII.
MONITOR, AQM_ID, SET_SENSOR_NAMES, CHECKSUM
MONITOR, AQM_ID, SENSOR_1, NAME_1, SENSOR_2, NAME_2, SENSOR_3, NAME_3, SENSOR_4,
NAME_4, SENSOR_5, NAME_5,
SENSOR_6, NAME_6, SENSOR_7, NAME_7, SENSOR_8, NAME_8, SENSOR_9, NAME_9,
SENSOR_10, NAME_10,
SENSOR_11, NAME_11, SENSOR_12, NAME_12, SENSOR_13, NAME_13, SENSOR_14,
NAME_14, CHECKSUM
* SET_SENSOR_NAMES - 1 byte command to set sensor names, refer section 2 for its value.
* SENSOR_1 ~ SENSOR_14 - 1 byte Specified sensor type has been used for the monitor.
* NAME_1 ~ NAME_14 - are the names of sensors. They are in ASCII format 7 char each, less than 7
USE blank char instead.
18A. Acknowledge reply command for setting sensor names, 4 bytes:
SENSOR, AQM_ID, SET_SENSOR_NAMES, CHECKSUM
19. Set concentration valid decimal numbers for monitors with display.
This is 4 bytes command, followed with 31 bytes sensor code and decimal numbers.
MONITOR, AQM_ID, SET_DECIMAL_NUMS, CHECKSUM
MONITOR, AQM_ID, SENSOR_1, DECIMAL_1, SENSOR_2, DECIMAL_2, SENSOR_3, DECIMAL_3,
SENSOR_4, DECIMAL_4, SENSOR_5, DECIMAL_5,
SENSOR_6, DECIMAL_6, SENSOR_7, DECIMAL_7, SENSOR_8, DECIMAL_8, SENSOR_9,
DECIMAL_9, SENSOR_10, DECIMAL_10,
SENSOR_11, DECIMAL_11, SENSOR_12, DECIMAL_12, SENSOR_13, DECIMAL_13, SENSOR_14,
DECIMAL_14, CHECKSUM
* SET_DECIMAL_NUMS - 1 byte command to set valid fraction digits for all the sensor readings.
* SENSOR_1 ~ SENSOR_14, - 1 byte each, sensor type code has been used for the monitor.
* DECIMAL_1 ~ DECIMAL_14 - 1 byte each, which is the values of decimal numbers for the sensor valid
reading.
19A. Acknowledge reply command for setting valid decimal numbers, 4 bytes:
SENSOR, AQM_ID, SET_DECIMAL_NUMS, CHECKSUM
20. AQM Monitor zero calibration status information request command, 4 bytes:
MONITOR, AQM_ID, ZERO_INFO, CHECKSUM
Example: 0x55 0x01 0xFC 0xAE
* ZERO_INFO - 1 byte zero calibration status information request command.
21. AQM monitor zero calibration status information reply command, 4 bytes:
SENSOR, AQM_ID, ZERO_INFO, ZERO_STATUS, CHECKSUM