IQM 60 V5 Pump Revision October 2010
31
SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12, SENSOR_13, SENSOR_14,
AQM_STATUS, CHECKSUM
* AQM_ID - 1 byte value, indicate the AQM ID number, which range is 1 - 255.
* SENSOR_NUM - 1 byte, specify the valid numbers of sensor has been used by the AQM box.
* SENSOR_1, SENSOR_2, SENSOR_3, SENSOR_4, SENSOR_5, SENSOR_6, SENSOR_7,
SENSOR_8,
SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12, SENSOR_13, SENSOR_14, - Specifies the
sensor type has been
used for the AQM box. The value will be any of Sensor type value.
* AQM_STATUS - 1 byte, used specify AQM auto report data and auto zero calibration, see section 2 for
details.
Example:
0x55 01 19 91
Followed by:
0x55 01 05 30 40 50 65 B0 00 00 00 00 00 00 00 00 00 03 CD
3A. Configuring AQM property acknowledgement command, 4 bytes:
SENSOR, AQM_ID, SET_CONFIGURATION, CHECKSUM
4. Request AQM Monitor property configuration command, 4 bytes,
MONITOR, AQM_ID, GET_CONFIGURATION, CHECKSUM
* GET_CONFIGURATION - 1 byte configuration request command, its value please see section 2.
Example:
0x55 0x01 0x18 0x92
5. Request AQM Monitor configuration command reply, 19 bytes (since V5.0):
SENSOR, AQM_ID, SENSOR_NUM, SENSOR_1, SENSOR_2, SENSOR_3, SENSOR_4, SENSOR_5,
SENSOR_6, SENSOR_7, SENSOR_8,
SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12, SENSOR_13, SENSOR_14,
AQM_STATUS, CHECKSUM
* AQM_ID - 1 byte value, indicate the AQM ID number, which range is 1 - 255.
* SENSOR_NUM - 1 byte, specify the valid numbers of sensor has been used by the AQM box, max
number is 14.
* SENSOR_1, SENSOR_2, SENSOR_3, SENSOR_4, SENSOR_5, SENSOR_6, SENSOR_7,
SENSOR_8, SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12, SENSOR_13, SENSOR_14, -
Specifies the sensor type has been
used for the AQM box. The value will be any of Sensor type value.
* AQM_STATUS - 1 byte, used specify AQM auto report data and auto zero calibration, see section 2 for
details.
6. Set monitor operation settings command 4 bytes followed by 8 bytes,
MONITOR, AQM_ID, SET_OPERATION, CHECKSUM
MONITOR, AQM_ID, SET_OPERATION, AUTO_REPORT_RATE, ZERO_CAL_INTERVAL,
AUTO_ZERO_READ_FREQ, OPRATION_STATUS, CHECKSUM
* SET_OPERATION - 1 byte command, see section 2 for values.
* AUTO_REPORT_RATE - 1 byte, specify the auto data report rate in minutes maximum 255 minutes. It's
only valid, when AQM_STATUS bit 0 is 1.
* ZERO_CAL_INTERVAL - 1 byte, specify the auto zero calibration interval in hours maximum 255 hours.
It's only valid, when AQM_STATUS bit 1 is 1.
* AUTO_ZERO_READ_FREQ - 1 byte, specify the auto zero reading frequency in hours, max 255 hours.
It's only valid, when AQM_STATUS bit 2 is 1.
* OPERATION_STATUS - 1 byte opration status, see section 2 for details.
6A. Acknowledge reply command of set operions, 4 bytes: