Page | 16
Commissioning your AQY 1
For most users, simply turning on the AQY 1 is enough. Data will start flowing.
The AQY 1 comes factory calibrated and will report measured air quality
parameters accurately.
For some users, with high data quality objectives, calibration against a ‘reference’
station is the best way to ensure highest correlation between AQY 1 and
reference. To do this install the AQY 1 at the reference station (within 10m of the
reference station inlet) allow it to run for minimum 72 hours then contact us for
further instructions (
).
Looking after your AQY 1
The sensors supplied with your AQY 1 are intended to provide trouble free
operation during the useful life of its sensors subject to geographical location.
WHO annual
average PM
10
(ug/m
3
)
Example Cities
Useful AQY 1 life
(months)*
0 - 50
Sydney, Los Angeles, London, Hong
Kong, Berlin, Singapore, Sao Paulo,
Tokyo, Moscow
10 - 14
51 - 100
Seoul, Bangkok, Istanbul, Shanghai,
Johannesburg
8 – 12
101 – 150
Beijing, Medina, Kuwait, Lagos
6 – 10
151 - 200
Delhi, Mumbai,
4 – 8
*Useful life means the time from installation until the point when the sensor can no longer be
remotely calibrated.
Calibration
Calibration parameters for the NO
2
, O
3
and PM
2.5
/PM
10
sensors are stored on
the associated sensor board so the sensors and connected board need to be
replaced with factory calibrated sensors for either calibration or in the case of
sensor failure.
To replace the sensors refer to the next section.