Aeroneb
®
Solo System Instruction Manual
21
Recharging the Battery
To recharge the battery, connect the AC/DC adapter to the
control module and connect to AC power source. The battery
status indicator is amber while charging and green when
fully charged. Allow a minimum of eight hours for the internal
battery to fully recharge.
Note:
If the control module is placed in long-term storage, it is
recommended that the battery be recharged every 3 months.
Installation for use with a ventilator
1.
For
adult breathing circuits
, connect the nebulizer
unit with adult T-piece into the inspiratory limb of the
breathing circuit before the patient Y (Figure 7).
For
pediatric breathing circuits
, connect the nebulizer
unit with pediatric T-piece into the inspiratory limb of the
breathing circuit before the patient Y (Figure 8).
For
neonatal breathing circuits
, connect the nebulizer
unit with the pediatric T-peice and the neonate adapters
approximately 30 cm (12 in.) back from the patient Y
(Figure 9), or neonatal set-up as shown in (Figure 10).
WARNING:
.
Condensate can collect and occlude ventilator
circuits. Always position ventilator circuits so that fluid
condensate drains away from the patient.
.
Always connect a bacteria filter to the expiratory inlet of
the ventilator. Otherwise the function of the expiratory
channel may be degraded.
Summary of Contents for Solo System
Page 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Page 2: ......
Page 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Page 4: ...This page has been intentionally left blank...
Page 46: ...This page has been intentionally left blank...
Page 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Page 48: ...This page has been intentionally left blank...
Page 55: ......