4.5

 

Fixed Head Mode Operation

   

 
 
 
 
 
 
 
 
 

 

4.5.1

   

The Fixed Head Modes (Figure 15) allow the operator to treat an enclosed area using one or 

 

 

both the two fixed heads located on top of the 3110 unit.  This mode works best when HVAC 

 

 

and ventilation have been isolated.  

 

4.5.2

   

Prior to activating the fixed head mode, position the 3110 unit in the treatment area. When 

 

 

using 1 Fixed Head Mode, the ideal unit location is where the end of the aerosol spray 

 

 

 

generated from the unit is in the center of the treatment area. When using 2 Fixed Heads 

 

 

 

Mode, the ideal unit location is where the unit is placed in the center of the treatment area.

 

4.5.3

   

Once the unit has been placed in the treatment area, plug the unit into power.

 

4.5.4

   

Next, Flip the red arming switch that is located to the right of the computer controller from the 

 

 

OFF position (Figure 10) to the ON position (Figure 11). This engages the unit into normal 

 

 

operation mode.

 

 

FIGURE 15

 

Summary of Contents for RDS 3110

Page 1: ...t for any purpose other than to evaluate this information This restriction does not limit the customer s right to use information contained in this document if it is obtained from another source without restriction The data subject to this restriction are contained in all sheets of this document This technical data is controlled under the Export Administration Regulations and may not be exported t...

Page 2: ...or Qualification 5 3 0 PRODUCT DESCRIPTION 6 3 1 Technical Specifications 6 4 0 SET UP OPERATION 6 4 1 Decontamination Set up 6 4 2 Computer Controller Settings Logic 7 4 3 Ambulance Mode Operation with ADP PT or ADP EX 12 4 4 APA Mode Operation 15 4 5 Fixed Head Mode Operation 16 4 6 Remote Head Only Mode Operation Tripod Mounted APA 20 4 7 Fixed Remote Head Mode Operation 22 5 0 PREVENTATIVE MAI...

Page 3: ...he solution is aerosolized and applied evenly to the treatment area A typical injection phase can last anywhere from 5 30 minutes depending on the size of the treatment area Dwell Phase Once an injection phase has completed the treatment area is allowed to sit for a minimum of 10 minutes in a dwell period The dwell period gives the aerosolized solution an opportunity to evenly distribute throughou...

Page 4: ...read and understand this document b RDS 3110 Operator Training individuals who will be operating the RDS 3110 unit for facility or asset decontamination using AeroClave solution must be trained and qualified See the requirements in Section 2 2 Operator Qualifica tions 2 1 3 Only a authorized AeroClave service technician may repair or maintain this equipment 2 1 4 The RDS 3110 unit is designed to o...

Page 5: ...approved solutions for decontamination Use of other solutions poses risk of injury machine failure and or unintended results and is prohibited 2 1 13 Use only de ionized or distilled water to flush demonstrate or practice with the system 2 1 14 Prior to transport or storage of the machine remove solution tank and transport or store separate from machine 2 1 15 Read and understand the AeroClave dis...

Page 6: ...egress paths door jambs windows bathroom vents hoods etc c Locations of equipment and furniture 4 1 2 Determine personnel exposure controls Controls must be in place to prevent unintentional personnel exposures Controls can be any or a combination of a Administrative where non decontamination personnel are removed from the area Shutting down the HVAC system to the area if possible is the preferred...

Page 7: ...nd materials needed in Section 1 3 4 1 8 The area is prepared All drawers cabinet doors and bathroom doors will be opened 4 1 9 The operator puts exposure controls and ventilation approach in place 4 2 Computer Controller Settings Logic 4 2 1 The RDS 3110 unit settings are prepared using a computer controller interface The interface is located on the underside of the lid of the RDS 3110 case 4 2 2...

Page 8: ... disclosure of data contained on this sheet is subject to the restriction on the cover page of this User s Manual 4 2 3 The RDS 3110 can be operated in seven modes The MODE button cycles through them 1 Port Mode Figure 2 FIGURE 2 2 Hand Applicator Mode Figure 3 FIGURE 3 ...

Page 9: ...anual Released July 1 2020 Page 9 Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User s Manual 3 1 Fixed Head Mode Figure 4 FIGURE 4 4 2 Fixed Heads Mode Figure 5 FIGURE 5 ...

Page 10: ... Released July 1 2020 Page 10 Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User s Manual 5 Tripod Only Mode Figure 6 FIGURE 6 6 1 Fixed Head and Tripod Mode Figure 7 FIGURE 7 ...

Page 11: ...10 Operator Manual Released July 1 2020 Page 11 Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User s Manual 7 2 Fixed Heads and Tripod Mode Figure 8 FIGURE 8 ...

Page 12: ...n ADP EX 4 3 2 Turn off the HVAC and ventilation in the treatment area 4 3 3 Open all interior compartments and close any exterior doors and pass throughs 4 3 4 When using an ADP PT Plug the attached ADP PT connection hose to the SmartPort located on top of the RDS 3110 unit When using an ADP EX Plug one end of the included ADP AS two way connection hose into the ADP EX port on the exterior of the...

Page 13: ...ges the unit into normal operation mode 4 3 6 Input the desired dosage time for the decontamination cycle The required dosage time for a typical ambulance is 6 minutes NOTE 6 minutes is the preset default dosage time on the RDS 3110 unit 4 3 7 Using the included handheld remote Figure 12 press the button to start the decontamination cycle OR press the Start button on the Computer Controller Figure...

Page 14: ...ed the red indicator lights on the unit will go dark If possible ventilate the treatment area Do not reenter area until air is clear minimum 20 minutes to reentry 4 3 11 Upon reentering the treatment area take note of any excess fluid on surfaces Excess fluid may remain in the air and on some surfaces in treatment areas with high humidity If there is any fluid remaining on surfaces you may wipe it...

Page 15: ...ocated on the RDS 3110 unit To do this align the white dot on the APA connector with the white dot on the RDS 3110 SmartPort insert the APA connector and turn the connector clockwise until it locks into position 4 4 4 Hand Applicator Mode does NOT require the red arming switch to be flipped up in the ON position to engage the system Figure 14 left Flip the switch down into the OFF position to enga...

Page 16: ...solution on known highly contaminated surfaces such as door handles keyboards toilet seats etc 4 4 11 It is recommended that operators practice with distilled or deionized water to become familiar with the operation of the device and the characteristics of the aerosol spray prior to performing actual field decontamination activities 4 4 12 If there is any excess fluid remaining on surfaces you may...

Page 17: ...o activating the fixed head mode position the 3110 unit in the treatment area When using 1 Fixed Head Mode the ideal unit location is where the end of the aerosol spray generated from the unit is in the center of the treatment area When using 2 Fixed Heads Mode the ideal unit location is where the unit is placed in the center of the treatment area 4 5 3 Once the unit has been placed in the treatme...

Page 18: ...fogging modes Fixed Head Tripod modes use an internal timer to determine how much solution will dispensed Determine the size of the treatment area then reference the chart located on the RDS 3110 unit for required injection times Figure 16 4 5 6 Injection time must be input prior to starting a Fixed Head Mode cycle To increase the amount of time injected press UP Figure 17 FIGURE 16 FIGURE 17 ...

Page 19: ...t of time press DOWN Figure 18 4 5 8 Once time injected has been input the cycle is ready to begin Figure 19 4 5 9 Using the included handheld remote Figure 20 press the button to start the decontamination cycle OR press the Start button on the Computer Controller Figure 21 NOTE If the unit needs to be turned off prior to the end of the cycle press either button again to terminate the cycle FIGURE...

Page 20: ...e has been reached the red indicator lights on the unit will go dark If possible ventilate the treatment area Do not reenter area until air is clear minimum 20 minutes to reentry 4 5 13 Upon reentering the treatment area take note of any excess fluid on surfaces Excess fluid may remain in the air and on some surfaces in treatment areas with high humidity If there is any fluid remaining on surfaces...

Page 21: ...screen upon engaging the red arming switch This error code will clear once the APA is mounted onto the tripod 4 6 4 Mount APA onto tripod 4 6 5 Position unit and or the tripod in the enclosed area NOTE When decontaminating a small space there may not be enough room for both the RDS 3110 unit and the tripod If the RDS 3110 unit is in another area ensure that the APA hose is not kinked or pinched wh...

Page 22: ...red indicator lights on the unit will go dark If possible ventilate the treatment area Do not reenter area until air is clear minimum 20 minutes to reentry 4 6 12 Upon reentering the treatment area take note of any excess fluid on surfaces Excess fluid may remain in the air and on some surfaces in treatment areas with high humidity If there is any fluid remaining on surfaces you may wipe it up or ...

Page 23: ...ar on the screen upon engaging the red arming switch This error code will clear once the APA is mounted onto the tripod 4 7 4 Mount APA onto tripod 4 7 5 Position unit and or the tripod in the enclosed area NOTE When decontaminating a small space there may not be enough room for both the RDS 3110 unit and the tripod If the RDS 3110 unit is in another area ensure that the APA hose is not kinked or ...

Page 24: ...red indicator lights on the unit will go dark If possible ventilate the treatment area Do not reenter area until air is clear minimum 20 minutes to reentry 4 7 12 Upon reentering the treatment area take note of any excess fluid on surfaces Excess fluid may remain in the air and on some surfaces in treatment areas with high humidity If there is any fluid remaining on surfaces you may wipe it up or ...

Page 25: ... maintenance flush re prime the RDS 3110 with disinfectant 5 2 Maintenance Steps 5 2 1 Replace on board operating reservoir with the provided maintenance reservoir 5 2 2 Ensure that the maintenance reservoir is filled with distilled or deionized water only 5 2 3 Start the operation of the RDS 3110 unit like normal 5 2 4 Use the remote control or APA trigger to stop the unit after it has run for th...

Page 26: ...dor Avoid contact with eyes Keep out of reach of children HMIS Rating Health 0 Flammability 0 Reactivity 0 PPE None Potential Health Effects Eye Contact Eye contact may cause mild eye irritation with discomfort Skin Contact Does NOT cause skin irritation and the product is NOT skin sensitizer Inhalation Does NOT cause any respiratory irritation If consumer product accidentally contacts strong acid...

Page 27: ... comply with OSHA 1910 106 Do not transfer contents to bottles or other unlabeled containers Do not reuse empty containers Keep out of reach of children Incompatible materials Strong acids Special Packaging Materials None Section 8 Exposure Control Personal Protection Engineering Controls Use in adequately ventilated areas Personal Protective Equipment Eye Face Protection Not required for consumer...

Page 28: ...sal must be in accordance with applicable Federal State Provincial and Local regulations Section 14 Transport Information Shipping Information Not regulated by DOT IMO IMDG and IATA ICAO for ground air or ocean shipments Section 15 Regulatory Information U S Federal Regulations TSCA All components appear in TSCA Inventory OSHA Refer to Section 8 for exposure limits CERCLA SARA Hazard Category Sect...

Page 29: ...is section show how to access the RDS 3110T expandable memory card where the data logging information is stored NOTE UNPLUG THE RDS 3110T UNIT FROM POWER BEFORE PROCEEDING 6 2 1 Remove the four screws on the exterior corners of the RDS 3110T yellow case There are two on the front of the case and to on the back of the case Figure B 1 6 2 2 Unscrew the two screws that secure the communication cable ...

Page 30: ...ect to the restriction on the cover page of this User s Manual 6 2 3 Lift the RDS 3110T unit straight up out of the yellow case and place the unit on a flat surface Identify that the stainless steel cover is visible Figure B 3 6 2 4 Remove the two screws securing the stainless steel cover Figure B 4 FIGURE B 3 FIGURE B 4 ...

Page 31: ...of this User s Manual 6 2 5 Remove the cover revealing the IDEC Smart Relay controller and the RDS 3110T system s electronics Figure B 5 6 2 6 On the front of the IDEC Smart Relay lightly press in the front of the carriage housing where the microSD card is located Upon release the carriage will pop out revealing the microSD card Figure B 6 FIGURE B 5 FIGURE B 6 ...

Page 32: ...ed on this sheet is subject to the restriction on the cover page of this User s Manual 6 2 7 Lightly press in the microSD card to release it from the carriage housing Figure B 7 6 2 8 Remove the microSD card by lightly pulling it straight out of the carriage housing Figure B 8 FIGURE B 7 FIGURE B 8 ...

Page 33: ...r to fit into the standard SD card slot 6 2 10 Once inserted into the computer the data log file will automatically download to the computer in a CSV comma separated values file format 6 2 11 After the file has been downloaded replace the microSD card into the IDEC Smart Relay carriage housing Note that the gold contacts on the underside of the microSD card should be facing upwards Figure B 9 6 2 ...

Page 34: ...the restriction on the cover page of this User s Manual 6 2 13 Lightly press in the front of the carriage housing back into the IDEC Smart Relay It will click when secured Figure B 11 6 2 14 Replace the stainless steel cover and re secure it using the two screws that were removed in Step 6 2 4 Figure B 12 FIGURE B 11 FIGURE B 12 ...

Page 35: ...r s Manual 6 2 15 Reinsert the RDS 3110T unit back into the yellow case and reconnect the communication cable onto the front of control panel securing it with the two screws Figure B 13 6 2 16 Re secure the RDS 3110T unit into the yellow case by replacing the four exterior screws removed in Step 6 2 1 Two on the front and two on the back Figure B 14 FIGURE B 13 FIGURE B 14 ...

Reviews: