RDS 3110 Operator Manual
Released July 1, 2020
Page 12
Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User’s Manual.
4.3
Port Mode Operation (with ADP
-
PT or ADP
-
EX)
4.3.1
The Port Mode (Figure 9, right) allows the operator to treat the enclosed space of a vehicle or
a room using either an ADP
-
PT or ADP
-
EX.
NOTE:
The ADP
-
PT and ADP
-
EX are optional components that must be purchased separately
and installed in the by an authorized AeroClave installer. An ADP
-
AS two
-
way connection
hose is necessary when using an ADP
-
EX.
4.3.2
Turn off the HVAC and ventilation in the treatment area.
4.3.3
Open all interior compartments and close any exterior doors and pass
-
throughs.
4.3.4
When using an ADP
-
PT:
Plug the attached ADP
-
PT connection hose to the SmartPort located
on top
of the RDS 3110 unit.
When using an ADP
-
EX:
Plug one end of the included ADP
-
AS two
-
way connection hose
into the ADP
-
EX port on the exterior of the ambulance and then plug the other end into the
SmartPort located on top of the RDS 3110 unit.
NOTE:
The error code “PLUG IN PORT CONNECTOR” (Figure 9, left) will appear on the
screen upon engaging the red arming switch. This error code is shown as a reminder to the
operator that an ADP Port must be connected prior to normal operation.
FIGURE 9
ADP
-
PT
ADP
-
EX (left) and ADP
-
AS (right)