3-Phase 3-Wire
∆
Balanced
(1 current probe) (3P3W
∆
b)
• VE/GND test lead to ground
• V1 test lead to L1
• V2 test lead to L2
• I3 probe to L3
3-Phase 4-Wire Y
(3P-4WY)
• VE/GND test lead to ground
• VN test lead to neutral (N)
• V1 test lead to L1
• V2 test lead to L2
• V3 test lead to L3
• IN probe to neutral
• I1 probe to L1
• I2 probe to L2
• I3 probe to L3
3-Phase 4-Wire Y Balanced (3P-4WYb)
• VE/GND test lead to ground
• VN test lead to neutral (N)
• V1 test lead to L1
• IN probe to neutral
• I1 probe to L1
3-Phase 4-Wire Y 2½ Element
(3P-4WY2)
• VE/GND test lead to ground
• VN test lead to neutral (N)
• V1 test lead to L1
• V3 test lead to L3
• IN probe to neutral
• I1 probe to L1
• I2 probe to L2
• I3 probe to L3
VN
V3
Source
Load
V2
V1
VE/GND
IN
I3
I2
I1
L1
L2
L3
N
L2
L3
L1
N
VN
V3
Source
Load
V2
V1
VE/GND
IN
I3
I2
I1
L1
L2
L3
N
L2
L3
L1
N
VN
V3
Source
Load
V2
V1
VE/GND
IN
I3
I2
I1
L1
L2
L3
L2
L1
L3
VN
V3
Source
Load
V2
V1
VE/GND
IN
I3
I2
I1
L1
L2
L3
N
L2
L3
L1
N