background image

•  Test current

 is the current to be used when running a test. Available 

settings are 50, 100, 150, or 200A. You can also select Manual to use the 

rotary knob on the instrument to control the current level during the test.

•  Language

 specifies the language to be used by the instrument’s display.

•  Auto Save

 specifies whether or not a test is automatically saved at the time 

it is performed. 

• 

Fan

 defines whether the fan is always on or runs in Auto mode (only turns 

on when the instrument needs to cool its internal temperature). 

•  Date format

 selects the format of dates to be used by the instrument 

(European or USA). 

•  Time format

 specifies whether the time is to be displayed using a 12 hour 

(AM/PM) or a 24 hour clock. 

On the right side of the Configuration Panel dialog box are five buttons:

•  Read from Instrument 

resets the options in this dialog box to reflect the 

settings as they are set in the instrument.

•  Write to Instrument 

writes the configuration settings to the instrument.

•  Set Clock 

displays the Date/Time dialog box. This enables you to set the 

instrument’s time to a specific date and time, or synchronize the instrument’s 

clock with the PC’s clock.

•  Delete Tests 

erases the instrument’s memory. Note that after you confirm 

that you want to proceed with the deletion; all objects are immediately 

removed from memory, irrespective of whether you save any other changes 

made through the Configuration Panel dialog box.

•  Cancel 

closes this dialog box discarding any changes specified without 

writing them to the attached instrument.

Downloading Stored Tests

1.

  To download stored tests, select 

Instrument 

from the menu bar, then click 

Download

. If a connection had not been previously established, the Connec-

tion dialog box will be displayed. Otherwise, the Select Tests dialog box will 

appear displaying a list of tests stored in the instrument.

2.

  Select the test you want to download by clicking it. You can select multiple 

tests by holding down the 

Ctrl

 key while clicking your selections. Click 

Select 

All

 to select all listed tests to download. After you have finished making selec

-

tions, click the 

Download

 button.

3.

  To edit the name and other properties of the test, double-click on the test’s 

name on the left side of the Control Panel.

4.

  Click on the desired test to download (hold the Ctrl key to select multiple 

tests), then click the 

Download 

button.

5.

  To edit the name and other properties of the test, double-click on the Test’s 

name on the left side of the Control Panel.

 

 

NOTE:

 

For complete operating instructions and more information on using 

DataView, refer to the product user manual that is supplied on the USB stick.

Summary of Contents for Model 6292

Page 1: ...Micro Ohmmeter Model 6292 CHAUVIN ARNOUX GROUP www aemc com ENGLISH Quick Start Guide...

Page 2: ...the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For...

Page 3: ...th the instrument Micro Ohmmeter Model 6292 Cat 2129 83 Set of Two 25 ft Kelvin Clips 200A Hippo Cat 2129 72 Large Classic Tool Bag Cat 2133 71 Green Ground Lead w Clamp Cat 2129 88 USB Cable 5 ft Cat...

Page 4: ...nt is protected by double insulation Battery Important instructions to read and to fully understand Useful information or tip to read The CE marking guarantees conformity with European directives and...

Page 5: ...C terminal Ensure the terminals are free of any voltage in relation to earth ground point and between each other Take into account that in a substation you will find in disconnected points high potent...

Page 6: ...ector 8 LCD Display 9 Fuse 10 Power cord connector 11 Fan 12 Alphanumeric keypad and function buttons 13 Rotary test knob 14 USB communication port 15 ON OFF switch Current Probe Connector 1 2 3 4 5 1...

Page 7: ...e rotary knob has two primary purposes 1 adjust the current during a test and 2 navigate through menus and select options When the main screen is displayed a short press will perform the MENU button f...

Page 8: ...the connection and try again If the problem persists contact technical assistance 3 Before turning the instrument on connect the test probes to the device to be tested and to the appropriate front pan...

Page 9: ...nnection to the instrument Then from the menu bar select Instru ment then click Configure The Configuration Panel dialog box will appear This dialog box allows you to configure options for each of the...

Page 10: ...you confirm that you want to proceed with the deletion all objects are immediately removed from memory irrespective of whether you save any other changes made through the Configuration Panel dialog b...

Page 11: ...MA 02035 USA Phone 800 343 1391 508 698 2115 Fax 508 698 2118 E mail techsupport aemc com NOTE Do not ship instruments to our Foxborough MA address Limited Warranty The Model 6292 is warranted to the...

Page 12: ...08 17 99 MAN 100394 v7 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Reviews: