6
Digital Thermometer Model K2000F and Temperature Probe Model ST2-2000
Common Mode Rejection Rate:
Up to 380V (50/60Hz) at thermocouple. No effect on the measurement.
Dielectric Strength:
4000V AC applied between thermocouple connector and the case.
Output:
1mV DC/°F
Input:
Standard type K (subminiature SMP connector)
Power Supply:
9V battery
Battery Life:
15,000 - 10s measurements; red LED for low battery
Dimensions:
6.1 x 2.09 x 1.18” (150 x 53 x 30mm)
Weight:
(7 oz.) 200 g.
Lead:
5 ft. (1.5m) lead with shielded 4mm banana plugs
Thermocouple (supplied):
Stainless steel needle probe K-type Model SK1 (-60°F to 1850°F)
Linearity Correction Curve
This curve gives the value that is necessary to add to the reading, in order
to take variations due to non-linearity into account.
Example: Reading: +1100°F Linearity Correction: -5°F
Corrected Reading: 1100°F ± 1% ± 2°F = ± 13°F
1100°F - 5°F = 1095 ± 13°F
-10
-50
32
100
20
0
30
0
40
0
50
0
60
0
70
0
80
0
90
0
1000
1100
120
0
130
0
140
0
150
0
160
0
170
0
180
0
190
0
2000
-5
0
Deg. F
Dt in Deg.
F
5
10
15
20
25