Power Quality Analyzer PowerPad
®
IV Model 8345 Quick Start Guide
Precautions Before Use
This instrument complies with safety standard IEC/EN 61010-2-030 or BS EN
61010-2-030. The leads comply with IEC/EN 61010-031 or BS EN 61010-031.
The current sensors comply with IEC/EN 61010-2-032 or BS EN 61010-2-032
for up to 600 V in CAT IV.
Failure to observe the precautions for use may create a risk of electric shock,
fire, explosion, or destruction of the instrument and installations.
■
Only competent and accredited personnel may perform troubleshooting
or metrological checks
■
The operator and responsible authority must read and understand the
various precautions to take before and during use
■
The operator must have knowledge and awareness of electrical hazards
when using this instrument
Do not use the instrument in an unspecified manner; otherwise, the
protection that the instrument provides may become compromised and
endanger you
Do not use the instrument on networks that exceed the instrument’s
specifications for voltage or category
Do not use the instrument if it seems to be damaged, incomplete, or
improperly closed
Do not use the instrument without its battery
Before each use, check the condition of the insulation on the leads,
housing, and accessories. Any item with deteriorated insulation (even
partially) must be set aside for repair or scrapping
Ensure that your instrument is completely dry before use. If it is wet, you
must dry it completely before connecting or using it
Use only the supplied leads and accessories. If you use any leads or
accessories with lower voltage or category ratings, you are limited to
lowest voltage or category rating
Use personal protection equipment when appropriate
Keep your hands away from the instrument’s terminals
Keep your fingers behind the physical guards when handling the leads,
test probes, and alligator clips
Use only the manufacturer-provided power supply unit and battery pack
because these items have specific safety components
At hazardous voltages, certain current sensors must not be placed on or
removed from bare conductors. Please refer to each sensor’s data sheet
and comply with their handling instructions