12
Power Quality Analyzer PowerPad
®
IV Model 8345 Quick Start Guide
Measurement Configuration Buttons
Button Purpose
to access the configuration screen
to access the measurement configuration options
to specify the calculation methods
to specify the nominal frequency (50 Hz or 60 Hz), nominal voltage
(with neutral), and nominal voltage between phases
to choose the real-time values to display
to specify the waveform mode
to specify the reference of levels of harmonics and mains signalling
voltage
to specify the curve of maximum MSV voltages as a function of
frequency. There are 5 preset points that you can modify
to choose the connection of the instrument according to the
distribution network
to specify the voltage ratios, the current sensor ratios, and the range
of the sensor
to choose the ratios of phase-to-neutral voltages (with neutral)
to choose the ratios of phase-to-phase voltages (without neutral)
to choose the ratios and range of current sensors
to reverse the current sensors
to configure the Energy mode
to configure the Trend mode (choice of values to record)
to configure the Transient mode (choice of triggering levels)
to configure the Inrush Current mode (choice of triggering levels)
to configure the alarms (choice of thresholds)
to enter the Monitoring mode, which is configured using the
application software
Taking Screenshots
You have two ways to take a screenshot:
Hold the
button until the symbol in the status bar turns yellow
and then black . Then, release the
button
Press the symbol in the status bar at the top of the display. The
symbol in the status bar will turn yellow and then black