35
9.3.3 Examples
Command
Response
Notes
R*
F15000000W800000N80D101R1
0000X2000R10010X2100…R10
100X2050*
Freq = 15MHz, Width = 800KHz, 80
points (81 actual), Data format = RX
1
st
pt. 1.0K+j200; 2
nd
pt.
1.001K+j210 …
R
F15000000W800000N80D101R1
0000X2000R10010X2100…R10
100X2050*
Note that the data dump response
always has an EOS marker whether
* follows R command or not.
M01
M01S001F200000W4000A0A24
…A50000*
Read Save memory 1, Freq=200K,
Width=4KHz, Instrument mode =
VIA; Memory slots = 24, … Max
center freq=50MHz
S000
S002F500000W32000A1A5…A2
00000*
Bravo MRI II’s current settings:
Freq=500K, Width=32KHz,
Instrument mode = SWR; Freq
step=5KHz, … Max center
freq=200MHz
F5000000* *
Center freq = 5MHz
W800000*
*
Sweep width = 800KHz
D102
*
Set unit to return total Z and Z angle
Q*
Q345*
Q is 34.5
D*
D8514*
Distance is 8514 mm
Command
Response
Notes
S100F1000000….A660*
*
Send unit setup data. The
cable velocity factor is 0.660
and the optional read only
data was not sent.
F5000000W800000*
*
Center freq = 5MHz
Sweep width = 800KHz